GW Instek MDO-2302AG Mixed-Domain Oscilloscope

USER MANUAL - Page 84

For MDO-2302AG. Series: MDO-2000A Series

PDF File Manual, 339 pages, Read Online | Download pdf file

MDO-2302AG photo
Loading ...
Loading ...
Loading ...
MDO-2000A series User Manual
82
Progress Indicator
Indicates the number of segments that have to
been captured relative to the set number of
segments.
Run/Stop Indicator
Stop: The segments have finished
acquiring or have been stopped.
Run: The scope is ready to acquire
segments.
Set the Number of Segments
Note
Before the Segment function can be used, set the
trigger settings as appropriate for the signal you
wish to use. The number of segments that can be
used depends wholly on the record length. See
page 78 to set the record length.
Record length
Number of segments
1000 pt.
1 ~ 29000
10k pt.
1 ~ 2900
100k pt.
1 ~ 290
1M pt.
1 ~ 20
Panel Operation
1. Press the Acquire key.
Acquire
2. Press Segments on the bottom
menu.
3. Press Select Segments and set the
number of segments from the side
menu.
Num of Seg
1~29000 (record length
dependant)
Set to Maximum
Sets to the maximum number
Loading ...
Loading ...