FLUKE T3000 FC KIT T3000 FC Wireless Temperature Kit

User Manual for the Fluke T3000 FC Wireless Temperature Kit - Page 46

For T3000 FC KIT.

PDF File Manual, 46 pages, Read Online | Download pdf file

T3000 FC KIT photo
Loading ...
Loading ...
Loading ...
3000 FC
Users Manual
42
Input Characteristics
Function
Overload
Protection
Input Impedance
(nominal)
Common Mode
Rejection Ratio
(1 k unbalance)
Normal Mode Rejection
1100 V rms >10 M <100 pF >120 dB at dc, 50 Hz or 60 Hz >60 dB at 50 Hz or 60 Hz
1100 V rms >10 M <100 pF >60 dB, dc to 60 Hz
1100 V rms >10 M <100 pF >120 dB at dc, 50 Hz or 60 Hz >60 dB at 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage
Typical Short Circuit Current
To 6 M 50 M
/
1100 V rms <2.7 V dc <0.7 V dc <0.9 V dc <350 µA
/
1100 V rms <2.7 V dc 2.000 V dc <1.1 mA
Function Overload Protection
Overload
mA Fused, 44/100 A, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10 minutes rest minimum
MIN MAX Recording
Function Accuracy
DC Functions The specified accuracy of the measurement function ±12 counts for changes >350 ms in duration.
AC Functions The specified accuracy of the measurement function ±40 counts for changes >900 ms in duration.