3000 FC
Users Manual
6
MIN MAX AVG Record Mode
The MIN MAX AVG record mode records the minimum
and maximum input values, and calculates a running
average of all measurements. The Product beeps when a
new high or low is sensed.
Note
For dc functions, accuracy is the specified
accuracy of the measurement function,
±
12 counts for changes longer than 250 ms in
duration.
For ac functions, accuracy is the specified
accuracy of the measurement function
±
40 counts for changes longer than 900 ms in
duration.
To start a MIN MAX AVG record session:
1. Make sure the Product is set to the correct
measurement function and on the correct range.
Autorange is disabled while in a MIN MAX AVG
record session.
2. Push . and Max show at the top of the
display. The measurement in the display is the
maximum value measured. It changes only when a
new maximum value is sensed.
3. To pause MIN MAX AVG record, push .
shows on the display while record is paused.
Recorded values are not deleted. To continue the
record session, push .
4. To exit and erase the MIN, MAX, and AVG values,
push for 1 second or turn the rotary switch.
5. To see the other recorded values (minimum and
average), push . Each push of the button
shows a different recorded value. The value shown in
the display is identified with Max, Min, or Avg to the
right of the MIN MAX icon.
Note
Sleep mode is turned off in MIN MAX AVG
record mode.