Company Record Screen
9.3 Change Company Name 9
9-7
9.2.3.12.1 Calibration Passed Status
This field is used to declare the status that an item (Instrument, Loop or Test Instrument)
should change to if it passes calibration. When the passed calibration is entered, upon
saving, a message will appear asking if you would like the system to automatically
update the item’s status. A few examples of statuses that would make sense for this field
are “In Service” or “Active”.
9.2.3.12.2 Calibration Failed Status
This field is used to declare the status that an item (Instrument, Loop or Test Instrument)
should change to if it fails calibration. When the failed calibration is entered, upon
saving, a message will appear asking if you would like the system to automatically
update the item’s status. A few examples of statuses that would make sense for this field
are “Out of Service” or “Needs Repair”.
9.2.4 Default Reports Tab
This tab page provides the fields in which to assign default reports to this company that
will be used when printing directly from a particular screen in the program.
9.3 Change Company Name
DPCTrack2 gives you the opportunity to change the Company name. From the HOME
screen select TOOLS “Change Company Name”. You must be a member of the
ADMIN group to access this feature.