4.Advanced User Guidebook
46
than 20Hz, and the amplitude should be no less than 5mv. Otherwise, the Autoset
function may be invalid.
[Run/Stop] button
Enable or disable sampling on input signals.
Notice: When there is no sampling at STOP state, the vertical division and the
horizontal time base of the waveform still can be adjusted within a certain range, in
other words, the signal can be expanded in the horizontal or vertical direction.
When the horizontal time base is 50ms, the horizontal time base can be expanded
for 4 divisions downwards.
[Copy] button
This button is the shortcut for Save function in the Utility function menu. Pressing
this button is equal to the Save option in the Save menu. The waveform, configure or
the display screen could be saved according to the chosen type in the Save menu. For
more details, please see "How to Save and Recall a Waveform" on P28.