FLUKE 1503 Insulation Resistance Meter

User Manual for the Fluke 1503 Insulation Resistance Meter - Page 30

For 1503.

PDF File Manual, 36 pages, Read Online | Download pdf file

1503 photo
Loading ...
Loading ...
Loading ...
1507/1503
Users Manual
26
Model 1503
Output Voltage Display Range Resolution Test Current
Accuracy
± (% of Rdg + Digits)
500 V dc
(0 % to + 20 %)
0.01 to 20.00 MΩ 0.01 MΩ
1 mA @ 500 kΩ ± (2.0 % + 5)
20.0 to 200.0 MΩ 0.1 MΩ
200 to 500 MΩ 1 MΩ
1000 V dc
0 % to + 20 %)
0.1 to 200.0 MΩ 0.1 MΩ
1 mA @ 1 MΩ ± (2.0 % + 5)
200 to 2000 MΩ 1 MΩ
IEC 61557 Specification
The following tables are a requirement for European labeling.
Measurement Intrinsic Uncertainty Operating Uncertainty
[1]
Volts ± (2.0 % + 3) 30 %
Earth-Bond Resistance ± (1.5 % + 3) 30 %
Insulation Resistance
Depends on test voltage and range.
See Insulation Test specifications.
30 %
[1] This specification comes from the standard and indicates the maximum amount allowable by the standard.