MDO-2000A series User Manual
330
Probe Specifications
GTP-100B-4
Applicable to: MDO-2102A/AG
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 100MHz
Input Resistance
10Mwhen used with oscilloscopes
with 1M input.
Input Capacitance
14.5pF to 17.5pF
Compensation Range
5pF to 30pF
Max. Input Voltage
600V DC + ACpk
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 10MHz
Input Resistance
1M(oscilloscope input resistance)
Input Capacitance
85pF to 115pF
Max. Input Voltage
200V DC + ACpk
Operating Cond.
Temperature
10°C to 50°C
Relative Humidity
85%
GTP-200B-4
Applicable to: MDO-2202A/AG
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 200MHz
Input Resistance
10Mwhen used with oscilloscopes
with 1M input.
Input Capacitance
10.5pF to 17.5pF
Compensation Range
5pF to 30pF
Max. Input Voltage
600V DC + ACpk
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 10MHz
Input Resistance
1M(oscilloscope input resistance)
Input Capacitance
65pF to 105pF
Max. Input Voltage
200V DC + ACpk
Operating Cond.
Temperature
10°C to 50°C
Relative Humidity
85%