Siglent SDS1102CML+ Technologies Digital Storage Oscilloscope, 100 MHz

User Manual - Page 103

For SDS1102CML+. Also, The document are for others Siglent models: SDS1000DL+

PDF File Manual, 130 pages, Read Online | Download pdf file

SDS1102CML+ photo
Loading ...
Loading ...
Loading ...
SIGLENT
SDS1000CML+/SDS1000DL+ User Manual 89
Picture 2.13-7
2.13.3 Self Calibration
Self Calibration procedure can optimize the signal path at the most measurement
precision. You can run this procedure at any time. If the operating temperature
changes by more than 5° C or the units run more than thirty minutes, you should
do the self calibration.
When you run the self Calibration, you should cut off all the probes and lead.
Then press the “Utility” button to choose the “Do self cal” to show the self
Calibration Menu, and execute self calibration program according to the
prompting on the screen.
Picture 2.13-8