ITECH IT6302 30V 3A Triple Output DC Power Supply

Manual - Page 10

For IT6302.

PDF File Manual, 28 pages, Read Online | Download pdf file

IT6302 photo
Loading ...
Loading ...
Loading ...
IT6302 user manual
USER MANUAL
10
If the latest operation state of the power supply was lost, then the VFD will display (about
2S):
Syst Lost
Picture 6
During channel scan, the VFD will display:
Scan Chan
Picture 7
When checkout for one channel is failed, then the VFD will display (about 2S):
Scan Chan
Chan Lost
Picture 8
When the calibration data of one or more channel is lost, then the VFD will display (about
2S):
Scan Chan
Cal Lost
Picture 9
When the factory calibration data of one or more channel is lost, then the VFD will display
(about 2S):
Scan Chan
Fact Lost
Picture 10
1.2.4 Output verification
The following procedures verify that the power supply outputs the correct voltage and
current levels and properly responds to entries from the front panel.
Voltage Output Check
The following steps verify basic voltage function without load.
1) Press Power key to turn on the power supply
2) Press
On/Off
key to enable the output
Note: when key is gray, the power supply is in SET mode, then the VFD
displays the set voltage and current; when the key is lit, then the power supply is
in METER mode, the actual voltage and current display on the VFD. When the
output is OFF, the ---- will display at the current display area.
3) Set the voltage value
Adjust the voltage, then press to lit the key (indicates it is in the METER mode),
make sure that the set value and output value are same, and if the current displayed on
the VFD is nearly 0A.
4) Make sure the voltage can be adjusted from zero to the maximum rated value.
5) Check the other two channels by the same method.