Function introduction
EN
Copyright©Qingdao Hantek Electronics Co.,LTD HT360 series user manual
27
Figure5 Printing Paper
1 Printer: Press the power on key , and the indicator will flash blue.
2 Hardness tester: Main interface > Print > Connect, the instrument displays
"Connecting the printer...". At this time, the instrument cannot be operated. After about
one minute, when the printer successfully connects to the device, the printer blue light
is off, printing can be performed.
During printing of multiple measured values, press to stop the current printing.
Function
Description
Link State
Displays the connection status between the instrument and
the printer.
Curr Value
Print the current measured value. If the current measurement
is not completed, prompt "Measurement is not completed" and
no printing will be performed.
Group Set
Press to select one of the 10 groups.
Value Set
Select some values to print in some print groups.
Print All
Print all the values in all print groups selected
Table 5 Print menu functions
5.8 Firmware upgrade
1. Download and install the burning tool "DfuSe Demo v3.0.5". Download the upgrade
package (* * *. dfu).
2. In the shutdown state, insert the USB cable into the Type C port of the instrument,
and the instrument is connected to the computer. Keep holding down the key,
gently press the power key and release it. At this time, the indicator light in the lower
right corner starts flashing red and green, indicating that the instrument has entered the
Dfu mode. Release F1 key.