inaccurate test values, so it should be reinforced or supported at the back of the test
point;
3. The magnetism of the sample should be less than 30 Gauss.
4.5.1 Measurement
Before measurement, random hardness block can be used to inspect the
instrument, and its display error and repeatability shall not be greater than those
specified in Table 4.1.
1. Booting
a. Insert the impact device plug into the impact device socket on the upper side of the
instrument;
b. Press the key. When the power is on, the instrument enters the measurement
state.
2. Loading
Figure 4 Loading
Push the loading sleeve downward to lock the impact body; For DC-type impact
device, place the loading bar on the test surface, insert the DC-type impact device
into the loading bar until its stop position, and then the loading is completed.
Press the support ring of the impact device tightly on the surface of the sample, and the
impact direction should be perpendicular to the test surface.
3. Measuring
Press the release button on the upper part of the impact device to measure. At this
time, the sample, impact device and operator shall be stable, and the force
direction shall pass through the axis of the impact device;