SNA5000A Vector Network Analyzer User Manual 94
to the actual data, not the real measured data. By setting the value
discretization of the marker, the value of the marker can only be
taken at the sampling point, which ensures that the marker’s value
is from the measurement data and not interpolated data. To open
the value discretization mode, click the Discrete to select the On
option, at this time, the data of the marker is only selected from the
sampling points, when the coordinate value of the marker is not in
the sampling point, the marker is automatically moved to the nearest
sampling point. To turn off the value discretization, click the Discrete
to select the Off option.
3. Fixed marker: The fixed marker mode is closed by default. Fixed markers have
a fixed excitation position (x-axis coordinate), its response value (Y-axis
coordinate) will change according to the trace data update. To keep the
fixed value of the marker unchanged, click the Type command to select
the Fixed option, then the marker’s value will keep the current value
unchanged, and the marker does not move up or down with the trace
update. To restore the marker real-time update, click the Type command
again to select the Normal option.
4. Data format: The newly created marker uses the current trace’s data format by default.
To change the data format of the marker, click the Format and select the
corresponding data format in the pop-up menu. The following data formats are
available:
Trace Default
Log/Phase
Lin/Phase
Re/Im
R+jx
G+jB
Log Magnitude
Lin Magnitude
Phase
SWR
Delay
Real
Imaginary
Loading ...
Loading ...
Loading ...