SNA5000A Vector Network Analyzer User Manual 56
Isolation
Frequency response
transmission tracking
Power calibration types: Transmitter calibration (internal source power calibration) and
receiver calibration.
3.3 S parameter calibration
3.3.1 S parameter calibration Kit management
3.3.1.1 Overview
For most applications, the default calibration kit models provide sufficient accuracy for your
calibration. However, several situations may exist that would require you to create a custom
calibration kit:
Using a connector interface different from those used in the predefined calibration kit
models.
Using standards (or combinations of standards) that are different from the predefined
calibration kits. For example, using three offset SHORTs instead of an OPEN, SHORT,
and LOAD to perform a 1-port calibration.
Improving the accuracy of the models for predefined kits. When the model describes the
actual performance of the standard, the calibration is more accurate. For example, A 7
mm LOAD is determined to be 50.4 Ω instead of 50.0 Ω.
Modifying the THRU definition when performing a calibration for a non-insertable device.
Performing a TRL calibration.
Operation instructions:
Press “Cal” on the front panel, Cal > Cal >Cal Kit to enter the Manage Cal Kits page. You can
edit, import, and create or delete calibration kits.