SNA5000A Vector Network Analyzer User Manual 110
Figure 4-5 Statistical data interface
4.6 Limit test
Limit Line is a test form that can visualize test data and results. The limit test compares
the measured data with the defined limit and provides the qualified or unqualified judgment
information of each measured data point. Limit line and limit test are not suitable for smith
chart and polar coordinate display formats, and limit line and limit test functions are
automatically disabled for these display types.
Limit table setting
Click Math→Analysis→Limit and the pop-up dialog box is shown in the figure below.
Limit test, sound type, limit table, global pass/fail can be set in this interface.
Figure 4-6 Limit test interface
Limit test:
Limit test on: Check this box to compare the data with the limit and display "pass"
or "Fail" judgment result.
Limit line on: Check this box to display the limit on the screen.
Sound on fail: When this box is selected, when the point on the trace fails in the limit
test, a beep will sound.
Sound type:
Sound on fail: When this box is checked, the equipment will beep when the test fails.
Sound on pass: When this box is checked, the equipment will beep when the test
passes.