Tripp Lite SMART1200XLHG SmartPro 120V 1kVA 750W Medical-Grade Line-Interactive Tower UPS, 4 Outlets, Full Isolation, Expandable Runtime

Owner’s Manual - Page 13

For SMART1200XLHG. Also, The document are for others Tripp Lite models: OMNISMART350HG, SMART700HG

PDF File Manual, 48 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
13
Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer
or the user of this Medical-Grade UPS should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions CISPR 11 Group 1 The Medical-Grade UPS uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and unlikely to cause any interference
in nearby electronic equipment.
RF Emissions CISPR 11 SMART700HG: Class A
SMART1200XLHG: Class A
OMNISMART350HG: Class B
The Medical-Grade UPS is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer
or the user of this Medical-Grade UPS should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
+ / - 6kV contact
+ / - 8kV air
Complies Floors should be wood,
concrete or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-3
+ / - 2kV for power supply lines
+ / - 1kV for input/output lines
Complies Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+ / - 1kV differential mode
+ / - 2kV common mode
Complies Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for .5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 seconds
Complies Mains power quality should be
that of a typical commercial or
hospital environment.
Power Frequency (50/60 Hz)
Magnetic Field
IEC 61000-4-B
3 A/m Complies Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Loading ...
Loading ...
Loading ...