Protmex PT6252B Handheld USB Wind Speed Meter Digital Anemometer

Product's Documents

Below are documents related to this product, you can read online or download:

Operation Manual

This is the main product document for model PT6252B.

The file format is pdf, 17 pages, you can download this manual here .

background
Table of Contents
1.
Safety information
……1
1.1
Preparation
……1
1.2
Usage
……1
1.3
Marking
……1
1.4
Maintenance
……2
2.
Description
……2
2.1
Part name
……3
2.2
Button description
……5
2.3
Sign definition
……5
3.
Specification
…………7
3.1
General
……7
3.2
Technique data
……8
4.
Operating guidance
…………11
4.1
Power On/Off
…………11
4.2
Reading hold
…………11
4.3
Backlight function
…………11
4.4
Wind speed measurement
…………12
4.5
Wind speed unit selection
…………12
4.6
Area setting
…………12
4.7
Air volume measurement
…………13
4.8
Air volume unit selection
…………13
4.9
Temperature unit selection
…………14
4.10
Temperature mode selection
…………14
4.11
USB real time data uploading
…………14
4.12
Replace battery
…………15
5.
Attachments
…………16
background
1
Digital Anemometer
Operation Manual
Safety Information
1. Safety Information
Please read carefully the following safety information
before usage and maintain this anemometer while using it
according to operating guidance, otherwise, the anemometer
may be damaged. The anemometer will provide satisfactory
services to you if you use and protect it appropriately.
1.1 Preparation
1.1.1 Please check for damage during transportation after
receiving the anemometer.
1.1.2 If it should be stored and shipped under hard
conditions, please confirm that whether the meter is
damaged.
1.2 Usage
1.2.1 The meter should be used in the range of specified
ambient temperature and humidity.
1.2.2 If you notice any abnormality or failure, it should stop
using.
1.2.3 Dont store or use the meter under the conditions of
direct sunlight, high temperature and high humidity.
1.2.4 Don’t touch the fan blades with excessive force.
1.2.5 Don’t directly expose the blades in the hard light to
avoid reading error.
1.3 Marking
The mark indicates compliance with EMC
requirements.
Important Safety Information
background
2
Digital Anemometer
Operation Manual
Description
1.4 Maintenance
1.4.1 Repair or maintenance should be implemented by
trained personnel.
1.4.2 If there is dust on the fan blade, please blow it with
clean air or scrub gently with a damp cloth and mild
detergent.
1.4.3 Clean the meter with a damp cloth and mild detergent.
Dont use abrasive material or solvent.
1.4.4 The meter should be powered off when it is not in use.
1.4.5 The meter will consume small current, about A,
after shutdown. If the meter is not to be used for a long
period, batteries should be removed to prevent
damaging the meter.
2. Description
- This meter is a digital anemometer for measuring the
ambient temperature, humidity, dew point temperature,
wet bulb temperature, wind speed, and air volume.
- This meter is a portable, professional measuring
instrument with large-screen LCD and backlight,
multi-unit switching functions.
- This meter can be used for hand-held or fixed
measurement.
background
3
Digital Anemometer
Operation Manual
Description
- This meter has the functions of reading hold, maximum,
minimum, etc.
- It has a low battery indicator and USB real time data
uploading functions.
2.1 Part name
Fan
LCD
FUN Function switching button
Power switch
UNIT Unit switching button
Maximum/Minimum button
HOLD/MODE hold display, switch temperature mode
/USBBacklight/USB real time data uploading
functions
background
4
Digital Anemometer
Operation Manual
Description
background
5
Digital Anemometer
Operation Manual
Description
2.2 Button description
Button:
Switch for meter powering on/off.
/USB button
Switch for turning on/off the backlight and USB
transmission.
HOLD/MODE button
Switch for reading retention and temperature mode.
FUN button
It is used for switching among the functions of wind
speed measurement, area setting and air volume
measurement. Long press for three seconds to enable or
disable "Auto Power-Off" function.
MAX/MIN button
Switch maximum/minimum/normal mode, long press to exit.
UNIT button
Switch unit, area (, ft²), long press to switch (°C , °F )
Wind speed (m/s, km/h, mil/h, ft/m, ft/s, knots),
Air volume (CMS, CMM, CFM).
2.3 Sign definition
Indicates Auto Power-Off status
H Reading hold state
VEL Wind speed measurement state
FLOW Air volume measurement state
AREA Area setting required by air volume
background
6
Digital Anemometer
Operation Manual
Description
DP Indicates that the current mode is dew point
temperature mode
WB Indicates that the current mode is wet bulb
temperature mode
°C, °F Temperature unit.
%RH Relative humidity sign
USB Indicates that USB real time data uploading is enabled
MAX Displays maximum after entering
maximum/minimum mode
MIN Displays minimum after entering
maximum/minimum mode
Indicates that the current area setting unit is square
meter
ft² Indicates that the current area setting unit is square foot
CMM Cubic meters per minute
CMS Cubic meters per second
CFM Cubic meters per minute
knots Nautical miles per hour, 1850 meters per hour
ft/s Feet per second
ft/m Feet per minute
m/s Meters per second
Km/h Kilometers per hour
mil/h Nautical miles per hour
Low battery indicator
background
7
Digital Anemometer
Operation Manual
Specification
3. Specification
The meter should be recalibrated under the condition
of 18°C28°C, relative humidity 75% every year.
3.1 General
3.1.1 Work height: Maximum 2000m
3.1.2 Work mode: Frequency of wind speed conversion
3.1.3 Display: LCD
3.1.4 Maximum show value: 9999
3.1.5 Sampling time: About 0.4s/time.
3.1.6 Low battery indicator:
sign displays on
LCD.
3.1.7 Work power: 9V 6F22 battery.
3.1.8 Operation environment:
Relative humidity 0~85%RH, no moisture
condensation
Temperature 0°C ~40°C , no moisture condensation
3.1.9 Detector (fan) operation environment:
Relative humidity 0~95%RH, no moisture
condensation.
Temperature -20°C~80°C, no moisture
condensation.
background
8
Digital Anemometer
Operation Manual
Specification
3.1.10 Storage environment:
Relative humidity 0~80%RH, no moisture
condensation
Temperature -10°C ~50°C , no moisture
condensation
3.1.11 Dimension: Meter 165LX85 WX38Hmm.
3.1.12 Weight: About 200g
3.2 Technique data
Ambient temperature: 23C, Relative humidity: <75%
3.2.1 m/s
Measuring range
Resolutio
n
Accuracy
0.80 ~ 30.00 m/s
0.01 m/s
±(2.0% reading + 50
characters)
30.00 ~ 40.00 m/s
For reference only
3.2.2 km/h
1.40~108.00 km/h
0.01km/h
±(2.0% reading + 50
characters)
108.0 ~ 144.0 km/h
For reference only
background
9
Digital Anemometer
Operation Manual
Specification
3.2.3 ft/s
1.30 ~ 98.50 ft/s
0.01 ft/s
±(2.0% reading +
50 characters)
98.50 ~ 131.20
ft/s
For reference only
3.2.4 knots
Measuring range
Resolution
Accuracy
0.80 ~ 58.30
knots
0.01
knots
±(2.0% reading
+ 50 characters)
58.30~77.70
knots
For reference only
3.2.5 mil/h
Measuring range
Resolution
Accuracy
0.90 ~ 67.20
mil/h
0.01mil/h
±(2.0% reading +
5 characters)
67.20~90.00
mil/h
For reference only
3.2.6 ft/m
Measuring range
Resolution
Accuracy
78 ~ 5900 ft/m
1ft/m
±(2.0% reading
+ 5 characters)
5900 ~ 7874 ft/m
For reference only
background
10
Digital Anemometer
Operation Manual
Specification
3.2.7 Air volume unit
CFM
0- 99990
(Area) 0 - 9.999 f
CMM
0- 99990
(Area) 0 - 9.999
CMS
0 - 9999
(Area) 0 - 9.999 m²
3.2.8 Ambient temperature, dew point temperature, wet bulb
temperature
Measuring range
Resolutio
n
Accuracy
-10°C 60°C
0.1°C
1.5°C
14°F ~+140°F
0.1°F
2.7°F
3.2.9 Relative humidity
Measuring range
Resolutio
n
Accuracy
(2080)%RH
0.1%RH
3%RH@25°C
(<20 or >80)%RH
0.1%RH
5%RH@25°C
background
11
Digital Anemometer
Operation Manual
Operating guidance
4. Operating guidance
4.1 Power On/Off
Press the key to turn on or off the anemometer
power.
4.2 Reading hold
In the measurement process, if the reading should be
kept, press "HOLD" key to lock the reading, and display the
HOLD symbol; press it again to unlock.
Note:
In the reading hold state, FUN, MAX/MIN and
UNIT keys are invalid.
4.3 Backlight
In the measurement process, if the ambient light is too dark
to read, you can press " " key to open the backlight.
Backlight timer is set to 15 seconds. During this period, you can
press " " key again to turn off backlight at any time.
Note:
The luminous body of backlight is LED with large
operating current. Frequently using backlight will
shorten battery life. Do not use backlight when
unnecessary.
When the battery voltage 7V, the " " (low battery)
symbol will show on the display. However, in the case
background
12
Digital Anemometer
Operation Manual
Operating guidance
of using the backlight, if the battery voltage ≥ 7V, the battery
voltage drops because of its larger operating current, "
"
symbol may show (when "
" symbol is showing, the
accuracy of the measurement can’t be guaranteed). At this
moment, you needn’t replace battery until the "
" symbol
display again under normal use condition without using
backlight.
4.4 Wind speed measurement
Place the detector (fan) into the test environment,
"VEL" symbol will display on the screen, measurement is
done with the fan surface perpendicularly to the wind
direction.
Note:
1. If the detector (fan) is not aiming at the wind direction,
which will bring the measurement error.
2. For steady wind, the detector (fan) will get maximum
reading when it is aiming at the wind direction.
4.5 Wind speed measurement
When using the meter to measure air volume, you can
press UNIT key to select the measurement unit you
required (m/s, km/h, mill/h, ft/m, ft/s, knots).
4.6 Area setting
To measure air volume, you should first determine the
area of air flue to be, area input steps are shown as
following:
background
13
Digital Anemometer
Operation Manual
Operating guidance
Press FUN key to make AREA display on the
screen.
Use the MAX/MINand UNIT keys to adjust value
and unit, after adjusting area unit (m², ft²), then press
MAX/MIN key. There should be an audible buzz,
indicating that area input is completed and settings are
saved.
To change the area setting, please repeat the step .
4.7 Air volume measurement
Place the detector (fan) into the test environment. Use
the FUN key to set the meter to the Air volume
measurement mode; the "FLOW" symbol should display on
the screen.
Measurement is done with the fan surface perpendicular
to the wind direction.
Note:
1. If the detector (fan) is not aligned in the wind direction,
the measurement can be skewed.
2. For steady wind, the detector (fan) will get maximum
reading when it is aiming at the wind direction.
4.8 Air volume unit selection
When using the meter to measure air volume, you can
press the UNIT key to select the measurement unit
background
14
Digital Anemometer
Operation Manual
Operating guidance
your required (CMS, CMM, CFM).
4.9 Temperature unit selection
Long press the UNIT key for 3 seconds to switch
temperature unit (°C , °F ).
4.10 Temperature mode selection
Long press the HOLD/MODE key for 3 seconds to
switch temperature mode (ambient temperature, dew point
temperature and wet bulb temperature).
4.11 USB real time data uploading
Long press the /USB key for 3 seconds to enable
or disable USB real time data uploading function.
Detailed steps are shown as follows:
1Use CD to install the PC software and the USB drivers.
2Power on the meter and connect it to a computer with
USB cable.
3Launch the PC software.
4Long press the /USBkey for 3 seconds to enable the
USB function of the meter.
5The PC software displays data and diagrams.
6Save or print data.
background
15
Digital Anemometer
Operation Manual
Operating guidance
4.12 Replacing battery
4.12.1 If the sign appears, this indicates that the
battery should be replaced.
4.12.2 Power off the meter and remove the battery cover.
4.12.3 Replace the old battery.
4.12.4 Install the battery cover properly.
background
16
Digital Anemometer
Operation Manual
Attachments
5. Attachments
Battery
9V, NEDA 1604, 6F22
1pcs
Packaging Bag
1pcs
Operation Manual
1pcs
USB wire
1pcs
(5)
CD
1pcs

Specifications

Protmex PT6252B Questions and Answers

Questions and Answers

Related Products