For ARG6PV456. Also, The document are for others Summit models: ARS1PV, ARS1PV456, ARS1ML, ARG1PV, ARG1PV456, ARG1ML, ARS3PV, ARS3PV456, ARS3ML, ARG3PV, ARG3PV456, ARG3ML, ARS6PV, ARS6PV456, ARS6ML, ARG6PV, ARG6ML, ARS8PV, ARS8PV456, ARS8ML, ARG8PV, ARG8PV456, ARG8ML, ARS12PV, ARS12PV456, ARS12ML, ARG12PV, ARG12PV456, ARG12ML, ARS15PV, ARS15PV456, ARS15ML, ARG15PV, ARG15PV456, ARG15ML, ARS18PV ARS18PV456 ARS18ML MLRS18MC ARS18MLMC R18AS HCRG19 ARG18PV ARG18PV456 ARG18ML MLRG18MC ARG18MLMC R18AG HCRG18 ARG31PVBIADA, ARG31PVBIADA456, ARG31MLBIADA, ARS32PVBIADA, ARS32PVBIADA456, ARS32MLBIADA, ARG61PVBIADA, ARG61PVBIADA456, ARG61MLBIADA, ARS62PVBIADA, ARS62PVBIADA456, ARS62MLBIADA

PDF File Manual, 44 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
17
CLEANING AND MAINTENANCE
Before cleaning, it is essential that you unplug the unit from the power line and transfer the contents
where they can be stored and monitored at the correct temperatures.
Wash the inner compartment with warm water and neutral detergent. DO NOT allow the control
panel, cables or plug to get wet. NEVER use corrosive detergents, wire brushes, or abrasive
scourers to clean your refrigerator. NEVER use metal or sharp implements to remove debris.
Dry all surfaces thoroughly.
To ensure trouble-free operation, the condenser should be cleaned every three months where
appropriate using a vacuum hose. The condenser is located at the back of the cabinet. In
exceptionally dusty locations, the condenser should be cleaned more often.
When transporting the unit, try to keep it vertical with angle of inclination between sides of the
refrigerator and the horizontal plane not less than 50o. Otherwise the compressor may be affected,