Summit ARG6PV456 6 Cu.Ft. ADA Height Vaccine Refrigerator

User Manual - Page 13

For ARG6PV456. Also, The document are for others Summit models: ARS1PV, ARS1PV456, ARS1ML, ARG1PV, ARG1PV456, ARG1ML, ARS3PV, ARS3PV456, ARS3ML, ARG3PV, ARG3PV456, ARG3ML, ARS6PV, ARS6PV456, ARS6ML, ARG6PV, ARG6ML, ARS8PV, ARS8PV456, ARS8ML, ARG8PV, ARG8PV456, ARG8ML, ARS12PV, ARS12PV456, ARS12ML, ARG12PV, ARG12PV456, ARG12ML, ARS15PV, ARS15PV456, ARS15ML, ARG15PV, ARG15PV456, ARG15ML, ARS18PV ARS18PV456 ARS18ML MLRS18MC ARS18MLMC R18AS HCRG19 ARG18PV ARG18PV456 ARG18ML MLRG18MC ARG18MLMC R18AG HCRG18 ARG31PVBIADA, ARG31PVBIADA456, ARG31MLBIADA, ARS32PVBIADA, ARS32PVBIADA456, ARS32MLBIADA, ARG61PVBIADA, ARG61PVBIADA456, ARG61MLBIADA, ARS62PVBIADA, ARS62PVBIADA456, ARS62MLBIADA

PDF File Manual, 44 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
13
F0
Defrost Type
00=defrost by
turning of comp.
00
C11
Comp.
Force
Running
Time
00=comp. Stop /
01-99min=starting
time
0 min
F1
Max defrost
duration
01~60 min.
20 min.
C12
Alarm
Output Type
00=contact
actuation when
alarm / 01=contact
disconnect when
alarm
01
F2
Defrost
Interval
XXhr/time
00~24 hr.
00
C13
Normal
State temp.
display type
00=liquid bottle
temp / 01=cabinet
air temp.
00
F4
Temp.
displaying on
screen when
defrost
00= normal Temp;
01=latest Temp.
02=Fixed Temp. ;
03=showing “dEF”
on screen
00
CF
Temperature
Unit
°C=Celsius /
°F=Fahrenheit
°C
C1
High Temp
Air Alarm
C2~20.0°C / 68.0°F
9.0°°C / 48.2°F
do1
00=do not alarm /
01=99
min=delayed alarm
02 min
C2
Low Temp Air
Alarm
-19.9°C / 3.8°F~C1
1.0°°C / 33.8°F
do2
Comp.
status when
door open
0=stop /
01=original status
0 1
do3
Light status
when door
open
00=start /
01=original status
0 1