Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specied below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
3 V 0.15 MHz to 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
3 V 0.15 MHz to 80
MHz 6 V in ISM and
amateur radio bands
between 0.15 MHz and
80 MHz
Portable and mobile RF communications
equipment should be used no closer to any part
of the device than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter. Recommended
separation distance
d = 1.2 P 150 KHz to 80 MHz
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 80 MHz to 2.7 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Radiated RF
IEC 61000-4-3
10V/m 80 MHz to 2.7
GHz
10V/m
Table 3