Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specied below. The customer or the user of device
should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ±4 kV, ±8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8kV,
± 15 kV air
Floors should be wood, concrete, or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s)
to earth
± 0.5kV, ± 1 kV line(s)
to lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Table 2