Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specied below. The customer or the user of device
should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ±4 kV, ±8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8kV,
± 15 kV air
Floors should be wood, concrete, or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s)
to earth
± 0.5kV, ± 1 kV line(s)
to lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Table 2