Pro'sKit Pro'sKit MT-1217 3 3/4 digit Auto Range Digital Multimeter

User’s Manual - Page 13

For MT-1217.

PDF File Manual, 28 pages, Read Online | Download pdf file

MT-1217 photo
Loading ...
Loading ...
Loading ...
• Overload Protection: 250V DC or AC Peak Value.
WARNING: DO NOT input any voltage at this range for
safety!
2-2-7. Capacity (F)
A) Rotate function switch to “
“setting.
B) Or rotate the Range to ““setting. Press “SELECT” button to
select the capacity measurement mode.
C) Insert the black test lead to “COM” terminal and red test lead
to “mA“ terminal;
D) Connect the tested capacity by the test leads to “COM”, “mA
“ input terminal, the screen will show capacitance parameter.
(The relative value measurement could be conducted by
pressing “REL” button.)
Caution:
1. Fully discharge the tested capacitor in case it damages the
meter.
2. When measuring in-line capacitor, the power should be turned
off and all capacitors should be discharged completely.
3. It takes about 30 seconds to input stable reading at 200uF
Range.
Range Accuracy Resolution
40nF ±(5.0%+30d) 10pF
400nF ±(3.5%+8d) 100pF
4μF ±(3.5%+8d) 1nF
20μF ±(3.5%+8d) 10nF
WARNING: DO NOT input any voltage at this range for
safety!
12
Loading ...
Loading ...
Loading ...