81
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
The SFT65 Multi-function Thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the SFT65 Multi-function Thermometer should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3Vrms
3 V/ m
Portable and mobile RF communications equipment should be used no closer to any
part of the SFT65 Multi-function Thermometer, including cables, than the recommen
-
ded separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, ama
-
teur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electroma-
gnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the SFT65 Multi-function Thermometer is used exceeds the applicable RF compliance level above, the
SFT65 Multi-function Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the SFT65 Multi-function Thermometer.
Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.