Binder ED 23 Heating Oven With Mechanical Control

User Manual - Page 45

For ED 23. Also, The document are for others Binder models: BD / BD-UL (E2), ED / ED-UL (E2), FD / FD-UL (E2)

PDF File Manual, 69 pages, Read Online | Download pdf file

ED 23 photo
Loading ...
Loading ...
Loading ...
BD / ED / FD (E2) 03/2019 page 45/68
Chamber size
FD 23
FD 53
FD 115
FD 240
Interior dimensions
Width
mm / inch
222 / 8.74
400 / 15.75
600 / 23.62
800 / 31.50
Height
mm / inch
330 / 12.99
400 / 15.75
480 / 18.90
600 / 23.62
Depth
mm / inch
277 / 10.91
330 / 12.99
400 / 15.75
500 / 19.69
Interior volume
l / cu.ft.
20 / 0.7
53 / 1.9
115 / 4.1
240 / 8.6
Steam space volume
l / cu.ft.
36 / 1.27
77 / 2.72
158 / 5.58
308 / 10.88
Racks
Quantity of racks (regular)
2
2
2
2
Quantity of racks (max.)
4
5
6
7
Load per rack
Kg / lbs
12 / 26
15 / 33
20 / 44
30 / 66
Permitted total load
Kg / lbs
25 / 55
40 / 88
50 / 110
70 / 155
Weight
Weight (empty)
Kg / lbs
28 / 62
44 / 97
62 / 137
96 / 212
Temperature data
Temperature range, 5 °C above ambient
up to
°C / °F 300 / 572 300 / 572 300 / 572 300 / 572
Temperature
fluctuation
at 70 °C / 158 °F
± K
--
0,1
0,1
0,1
at 150 °C / 302 °F
± K
0,3
0,3
0,3
0,3
at 300 °C / 572 °F
± K
--
0,2
0,3
0,4
Temperature
uniformity
(variation)
at 70 °C / 158 °F
± K
0.8
0.8
0.7
0.8
at 150 °C / 302 °F
± K
2.5
3.0
2.6
2.9
at 300 °C / 572 °F
± K
4.5
6.7
4.6
6.3
Heating up time
to 70 °C / 158 °F
minutes
10
7
20
22
to 150 °C / 302 °F
minutes
25
24
28
27
to 300 °C / 572 °F
minutes
59
60
75
66
Recovery time
after door was
open for 30 sec
at 70 °C / 158 °F
minutes
5
4
3
2
at 150 °C / 302 °F
minutes
6
5
5
6
at 300 °C / 572 °F
minutes
10
9
13
13
Ventilation data
Air change
at 70 °C / 158 °F
x/h
59
59
29
19
at 150 °C / 302 °F
x/h
64
64
32
20
at 300 °C / 572 °F
x/h
53
53
26
18
Electrical data
(model versions FD023-230V, FD053-230V, FD115-230V, FD240-230V)
IP system of protection acc. to EN 60529
IP
20
20
20
20
Nominal
voltage (±10%)
at 50 Hz power frequency
V
230
230
230
230
at 60 Hz power frequency
V
230
230
230
230
Current type
1N~
1N~
1N~
1N~
Nominal power
kW
0.8
1.20
1.60
2.70
Power plug
Grounded plug
Chamber fuse
5x20mm / 230V / time-lag M
Amp
10
external
10
external
10
external
16
external
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Loading ...
Loading ...
Loading ...