Binder ED 23 Heating Oven With Mechanical Control

User Manual - Page 42

For ED 23. Also, The document are for others Binder models: BD / BD-UL (E2), ED / ED-UL (E2), FD / FD-UL (E2)

PDF File Manual, 69 pages, Read Online | Download pdf file

ED 23 photo
Loading ...
Loading ...
Loading ...
BD / ED / FD (E2) 03/2019 page 42/68
Chamber size
BD 23
BD 53
BD 115
BD 240
BD 400
BD 720
Electrical data
(model versions BD023-230V, BD053-230V, BD115-230V, BD240-230V, BD400-230V, BD720-230V)
IP system of protection acc. to EN
60529
IP 20 20 20 20 20 20
Nominal
voltage
(±10 %)
at 50 Hz power frequency
V 230 230 230 230 230 230
at 60 Hz power frequency
V 230 230 230 230 230 230
Current type
1N~
1N~
1N~
1N~
1N~
1N~
Nominal power
kW
0.2
0.4
0.4
0.68
0.85
1.25
Chamber fuse 5 x 20 mm
230V / 10A / middle-time-lag (M)
Amp
10
external
10
external
10
external
10
external
10
external
10
external
Power plug
Grounded plug
Installation category acc. to IEC 61010-1
II
II
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
2
Different electrical data for BD-UL constructed for the USA and Canada (model versions BD023UL-
120V, BD053UL-120V, BD115UL-120V, BD240UL-120V, BD400UL-120V, BD720UL-120V)
Nominal voltage (±10 %)
at 60 Hz power frequency
V 115 115 115 115 115 115
Current type
1N~
1N~
1N~
1N~
1N~
1N~
Nominal current
Amp
1.8
2.6
3.5
5.9
7.4
10.9
Power plug
NEMA
5-15P
5-15P
5-15P
5-15P
5-15P
5-20P
Chamber fuse 6,3 x 32 mm 250V /
super-time-lag TT
Amp
12,5
external
12,5
external
12,5
external
12,5
external
12,5
external
16
external
Additional temperature fuse class 1
(DIN 12880:2007)
internal internal internal internal internal internal
Environment-specific data
Noise level (mean value)
dB (A)
< 45
< 45
< 45
< 45
< 45
< 45
Energy consumption at 37°C / 98.6°F
Wh/h
11
11
20
33
56
80
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10. Technical data is
determined in accordance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
With option interior socket: If electrical devices are connected and operating inside the
chamber, the temperature range may be modified due to heat emission.
12.5 ED technical data
Chamber size
ED 23
ED 53
ED 115
ED 240
ED 400
ED 720
Exterior dimensions
Width, net
mm
inch
435
17.13
635
25.00
835
32.87
1035
40.75
1235
48.62
1235
48.62
Height, gross (incl. feet/castors)
mm
inch
495
19.49
620
24.41
705
27.76
825
32.48
1025
40.35
1530
60.24
Depth
mm
inch
520
20.47
575
22.64
645
25.39
745
29.33
765
30.12
865
34.06
Depth, gross (incl. door handle and
exhaust duct)
mm
inch
625
24.61
680
26.77
750
29.53
850
33.46
870
34.25
970
38.19
Wall clearance rear
mm
inch
100
3.94
100
3.94
100
3.94
100
3.94
100
3.94
100
3.94