MK / MKT (E5) 08/2020 page 27/162
The maximum hose length with nominal width ½ " is 5m. If a longer hose is necessary, a larger diameter
hose must be used.
Protect both ends of the hose with two of the four supplied hose clamps. Before turning on the chamber,
check the connection for leaks.
Water consumption:
The average water demand is not helpful for the design of the water supply line, since the supply line must
provide sufficient dimensions for peak loads.
Peak values occur briefly (<5 min) when temperature rapidly decreases from + 180 ° C to a much lower
value. Also the activation of the condensation protection leads to high water consumption.
Max. water flow rate (peak value):
MK 56: approx. 0,6 m³/h
4.3 Installation of the voltage and frequency changer
(chambers with voltage and frequency changer)
The voltage and frequency changer is supplied packed separately together with the chamber.
CAUTION
Risk of injury and danger of damages by lifting heavy loads and by sliding or tilting
of the voltage and frequency changer in case of improper lifting.
Injuries; damage to the voltage and frequency changer.
Do NOT lift the voltage and frequency changer by hand.
Lift the voltage and frequency changer from the pallet using technical devices (fork
lifter). Set the fork lifter only from the rear in the middle of the chamber.
Alternatively, the voltage and frequency changer can also be lifted at the eyelets on
the top by means of a lifting crane or fork lifter
(a) Eyelets for lifting with a lifting crane or fork lifter
(b) Positions for a fork lifter
Figure 11: Positioning of aids for lifting the voltage
and frequency changer
(a)
(b)
(a)
(b)