TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSES
POSSIBLE SOLUTION
The Unit Cannot
Power On
Are the batteries depleted? Recharge or replace the batteries.
Are the batteries installed correctly? Insert the batteries observing the correct polarity.
Stimulation Weak
Or Cannot Feel
Any Stimulation
Electrodes are dried out or dirty. Replace with new electrodes.
Electrodes will not stick to skin. Clean treatment area of dirt and/or oily
substances (lotion).
Lead wires are old/worn/damaged Replace with new lead wires.
Stimulation Is
Uncomfortable
Intensity is too high. Decrease intensity.
Electrodes are too close together Reposition the electrodes to be a minimum of
1½" apart.
Electrode active area size is too small Replace electrodes with ones that have an active
area no less than 25.0cm
2
(5cm x 5cm).
May not be operating the device according to the manual. Please check the manual before use.
NOTE: If the unit does not operate after taking these measures, contact your nearest dealer.