What to do if …..
Problem Reason Solution
Heart rate
symbol
Appears in the measurement
condition and ashes when
pulse is detected.
Measurement in progress, remain quiet.
Low battery
indicator
Appears when the battery
voltage is excessively low or the
positions of the batteries are
incorrect.
Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
Measuring
error
Appears when the accurate
blood pressure and pulse could
not be obtained.
Press “start/stop button again and
remeasure.
Check if cu is wrapped according
to instructions.
Check that there are no kinks in tube.
Check that hand is relaxed.
Check if talking or moving
during measurement.
Check if posture is correct.
E1 shows The cu is not secure. Refasten the cu and then measure again.
E2 shows The cu is very tight. Refasten the cu and then measure again.
E 10 or
E 11 shows
The monitor detected motion,
talking or the pulse is too poor
while measuring.
Relax for a moment and then measure again.
E 20 shows The measurement process does
not detect the pulse signal.
Loosen the clothing on the arm and then
measure again.
E 21 shows Measure incorrectly. Relax for a moment and then measure again.
EE 3 - EE15 Error while measuring. Retake the measurement. If the problem
persists, contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.