User Manual Rossmax TG250 Thermometer

Documents for Rossmax TG250

The following documents are available:
User Manual Photos
TG250 photo
1. When the appears on the upper right corner of the LCD,
the battery is exhausted and needs replacement.
2. Pull the battery cover off in the direction shown below.
3. Pull out the battery with a sharp objects such as pen. Keep
exhausted battery from
reaching by the children.
4. Place a new 1.5V D.C.
button size battery type
SR41 or LR41 or equiva-
lent in the chamber with
positive side faced up and
negative side faced down.
5. Install the battery cover
back.
1. To push the On/Off button by the display window. The display will
read for 2 seconds and beep sound will be heard.
2. Releasing the on/off button, last measured temperature will be
shown for about 2 seconds. Then shows the self-test temperature
37.0°C(98.6°F) , and then Lo°C (Lo°F) displays.
3. Wipe or dip the probe with alcohol for disinfection. Position the probe
in oral, rectal or axillary site.
4. Once the degree sign °C (°F) on the display has stopped ashing (usu-
ally within 10 to 40 seconds), a Beep signal will sound for approx. 10
times, the measured temperature is shown on LCD window. The tem-
perature reading will not change after the unit is removed from testing
position. If the temperature <37.8°C(100.0°F ). the buzzer will alarm
“Bi---Bi---Bi---Bi---”. If the temperature ≥ 37.8°C (100.0°F). the buzzer
whose frequency is higher will alarm “Bi-Bi-Bi----Bi-Bi-Bi--” about 10
times rather than”Bi---Bi---Bi---Bi---”
5. The unit will automatically turn off in 10 minutes for energy-saving.
However, to prolong the life of the battery, it is suggested to press the
On/Off button to turn the unit off once the measurement is done.
6. Please dispose of any used battery and the thermometer according
to the local regulations.
Note: After step 2, if the room temperature is greater than 32.0°C
(89.6°F ), it will be displayed the room temperature.
Note: Usage of the probe cover may result in a 0.1°C(0.2°F) differ-
ence from actual temperature.
Note: To clean the probe before and after using the thermometer
to ensure an accurate reading and avoid cross contamination ap-
proximately.
www.rossmax.com
Introduction
How to Use
Caution
Temperature readings are available in the Fahrenheit or Celsius scale
(°C / °F; located in the right of LCD.) With the unit off, press and
hold the On/Off Button for approximately 4 seconds to change the
current setting.
ºC /ºF Switchable
Battery Replacement
Avoid strong electromagnetic interference such as microwave ovens
and cell phones.
Please be noticed to keep the probe dry before putting the ther-
mometer into the carry case.
The thermometer contains small parts (battery, etc.) that can be
swallowed by children. Therefore never leave the thermometer un-
attended to children.
The thermometer should be cleaned after measurement no matter
the probe cover has been installed or not.
Keep battery away from children. Harmful is swallowed.
Battery should not be charged or placed into extreme heat as it
may explode.
Do not use the thermometer in ear. Intended use is for oral, axillary
and rectal measurement.
Use of the probe cover may result in a 0.1ºC(0.2ºF) discrepancy from
actual temperature.
Remove battery from the thermometer when not in operation for
a long time.
If the unit is stored in an extreme environment before measurement,
please make sure it has been rstly adapted in the room tempera-
ture, so that the unit can be measured normally.
Please note that this is a home healthcare product only, and it is not
intended to serve as a substitute for the advice of a physician or
medical professional.
Do not use this device for diagnosis or treatment of any health prob-
lem on disease. Measurement results are for reference only. Contact
your physician if you have or suspect any medical problems. Do not
change your medications without the advice of your physician or
healthcare professional.
This device may not meet its performance specication if stored or
used outside temperature and humidity ranges specied in speci-
cations.
If this device is used according to the operations instruction, periodic
re-calibration is not required. If you still have questions, please send
the device to dealers.
Please do not dispose of the product in the household waste at the
end of its useful life. Disposal can take place at your local retailer or
at appropriate collection points provided in your country.
This device is for oral, axillary and rectal use only.
High, prolonged fever requires medical attention especially for
young children. Please contact your physician.
For save reason, during children’s temperature measurement,
please keep them from crying, walking, talking and any related
dangerous activities.
Do not bend, drop or twist the thermometer. It is not shock-
proof.
Do not boil probe for disinfect.
Do not store the unit under direct sunlight, at a high tempera-
ture, in high humidity or dust. Neither store the unit under the
sunlight, nor at the high temperature or humidity.
Do not disassemble. See BATTERY REPLACEMENT to replace
battery.
Instruction Manual
Model: TG250/TG380
EN
Tightly place probe in the patient’s armpit for about 30 seconds.
* Please ensure the close contact between the probe and the armpit
of the baby or children.
Axillary Use
Troubleshooting
Cleaning and Disinfection
1. Clean the unit by wiping it with a dry cloth and sterilize the probe
with ethyl alcohol.
2. Don’t let the unit contact any chemical thinner.
3. Do not soak the display in water.
Error message
Problem Solution
The system is
not functioning
properly
Unload the battery, wait for 1
mins and repower it.
If the message reappears, con-
tact the retailer for service.
(bi-bi...bi)x10
Temperature
taken is higher
than 42.9°C
(109.2°F)
Take a new temperature
measurement after 6 sec.
Temperature tak-
en is lower than
32°C (89.6°F)
Take a new temperature
measurement after 6 sec.
Dead battery:
Battery icon is
ashing, can’t be
measurable.
Suggest to replace the battery.
Display on the LCD
Place the probe well under the patient’s tongue. Conrm the patient to
keep his mouth closing for about 20 seconds.
Do not drink hot or cold uids, exercise, and smoke or per-
form other activities prior to a reading. These activities will raise
or lower temperature readings when compared to your normal,
average temperature.
Oral Use
Probe
Battery
Cover
ON/OFF Button
Display
The Digital Medical Thermometer provides quick, highly accurate
reading over the human body temperature range. Please read the
instructions completely rst.
The quality of the thermometer has been veried and conforms
to the provisions of the EC council directive 93/42/EEC (Medi-
cal Device Directive) Annex I essential requirements and applied
harmonized standards. EN 12470-3: 2000 /A1: 2009 Clinical
thermometers-Part 3: Performance of compact electrical ther-
mometers (non-predictive and predictive) with maximum device.
Attention: Consult the accompanying documents. Please
read this manual carefully before use. Please be
sure to keep this manual.
Always uses probe cover when taking rectal measurements.
1. Commonly used for young children when it is difcult to take an
oral or axillary temperature.
2. Lubricate with a water-soluble gel. Do not use petroleum jelly.
3. Gently insert the probe tip no more than 1.5-2 cm into the rec-
tum for about 10 seconds. Do not force the tip into the rectum if
resistance is encountered.
4. Disinfect the thermometer after use. Refer to “cleaning and
disinfection“section.
Rectal Use
M: Last measured temperature.
: When the appears on the upper
right corner of the LCD, the battery is ex-
hausted and needs replacement.
Loading ...