So-Low U40-18

Product's Documents

Below are documents related to this product, you can read online or download:
User Manual Specification Other Documents User Service
U40-18 photo

User Manual

This is the main product document for model U40-18. Additionally, the document applies to other So-Low models: PV40-18

The file format is pdf, 24 pages, you can download this manual here .

background
To reduce the risk of fire, electric shock or injury to persons using this freezer, read all
instructions and follow basic safety precautions before using the unit, including the following:
Do not modify the plug provided with the freezer. If it will not fit the outlet,
have a proper outlet installed by a qualified electrician.
Do not position equipment so it is difficult to disconnect from the power supply.
freezer must be at least 6” away from any wall or object on any side.
While under warranty, do not attempt to repair or replace any part of the
freezer for servicing without first contacting the So-Low Service Department.
SAVE THESE INSTRUCTIONS
background
MEANING OF ILLUSTRATED SYMBOLS…………………………………...
STARTING INSTRUCTIONS…………………………………………………..
PRE-INSTALLATION INFORMATION………………………………………...
FREEZER STORAGE PROCEDURE……………………………………..….
● ALARM BATTERY REPLACEMENT…...……………………………………..
CLEANING PROCEDURE……………………………………………………..
DEFROSTING PROCEDURE………………………………………………….
CLEANING AIR CONDENSER………………………………………………...
CHANGING SETPOINT / OPERATION TEMPERATURE………….………
● ALARM SYSTEM………………………………………………………………..
8
● DRY CONTACT RELAY………………………………………………………..
8
● CONTROL KEYS & DISPLAYS………………………………………………..
9
PROGRAM MENU NAVIGATION…………………………………………......
10
● CALIBRATION PROCEDURE………………………………………………....
11
ALARM DIFFERENTIAL……...………………………………………………...
12
CONTROL HYSTERESIS…...………………………………………………....
13
POWER STATUS ALARM INDICATOR………………………………...........
13
MAINTENANCE CHECKLIST
DIAGRAMS
background
Various symbols are used in this safety manual in order to use the unit without
danger of injury and damage of the unit. Be sure that you understand the
warnings and cautions in this manual before operating the unit.
background
background
background
background
background
background
background
background
TOUCH KEYS
FUNCTION
DESCRIPTION
Up Key
Press and release to increase the current control set point
(while in normal control mode) or to change the value of the
selected parameter (while in the Setup Menu).
Down Key
Press and release to decrease the current control set point
(while in normal control mode) or to change the value of the
selected parameter (while in the Setup Menu).
Scroll Key
Press and hold for at least 5 seconds to enter the Setup
Menu.
The upper display will show SET and then release the key.
Reset Key
Press and release to return the display to the home screen
(while in User Menu or Setup Menu).
PV
SV
-888
-888
R
R
°C
AL1
AL2
AL3
OUT1
MAN tune
COM
background
When the desired sub-menu parameter is selected,
THE FOLLOWING VALUES WERE FACTORY SET WHEN THIS
UNIT WAS SHIPPED. INCORRECT MODIFICATION OF THESE
VALUES CAN VOID THE WARRANTY OF THE UNIT.
PARAMETER
SUB PARAMETERS
DESCRIPTION
°F
°C
bASE
SP1L
Lower Setpoint
-88
-67
bASE
SP1H
Upper Setpoint
+32
0
bASE
UNIT
Temperature Scale
°F
°C
bASE
OFTL
Calibration
ALRM
A1DV
Alarm Differential
22
12
ALRM
A1HY
Alarm Hysteresis
0.1
0.1
OUT
O1HY
Output Hysteresis
3.6
2.0
R
background
CAUTION
INCORRECT MODIFICATION OF THESE VALUES CAN RESULT IN
MALFUNCTION OR UNEXPECTED OPERATION OF THE UNIT,
AND MAY VOID THE WARRANTY OF THE UNIT.
R
background
R
background
R
background
Q:
background
COMPRESSOR MODEL
HP
VOLTAGE
HERTZ
PHASE
PART #
TECUMSEH AJB2433ZXA
1
115
50/60
1
AJB24-115
TECUMSEH AJB2433ZXD
1
208/220/230
50/60
1
AJB24-208
EMBRACO FFI12HBX
1
/
3
115
50/60
1
FF12-115
DANFOSS SC15FTX
1
/
3
115
50/60
1
SC15-115
DANFOSS SC18FTX
1
/
2
208/220/230
50/60
1
SC15-208
TEMPERATURE CONTROL PARTS
PART #
FDC C42
FDC-C42-VOLTAGE
FDC nCOMPASS
nCOMPASS
ELECTRICAL PARTS
PART #
Heater Harness No. H-200
217-VOLTAGE
Refrigeration Switch No. 2X464
TOGGLE
Condenser Fan Motor No. GE-5411 - 115/60/1
500-115
Condenser Fan Motor No. GE-5421 - 230/50-60/1
500-VOLTAGE
Electrical Cord No. 8-3 (Please Specify Voltage)
PWRCRD-VOLTAGE
Control Board No. CECB2TUV (Please Specify Voltage)
231-VOLTAGE
REFRIGERATION HIGH STAGE PARTS
PART #
Air Cooled Condenser No. 3CZ0602B
254
Drier No. C-053-S
256H
Capillary Tube
HS-17, HS-20
Oil Separator, Temprite Series 900 (If Applicable)
900
REFRIGERATION LOW STAGE PARTS
PART #
Pressure Control No. 20PS01-0039
259
Receiver Condenser
RCN-LS
Drier No. CO-52S-S
256L
Capillary Tube
LS-28, LS-31
Oil Separator, Temprite Series 900 (If Applicable)
900
HARDWARE PARTS
PART #
Latch No. METL-L1-99
REX37L1-3
Chest Hinge
59-928M
Upright Hinge No. Polar 109-LH
59-928U
Cabinet Gasket
NX504B1
Lid or Door Gasket
PSOS
Grill No. 650H
356F, 356S
Sub-Lids (Must have Model Number)
SL-MODEL NUMBER
Inner Door (Must have Model & Serial Number)
357-MODEL NUMBER-SERIAL NUMBER
Clips & Rollers for Inner Doors (Quantity 10 minimum)
405
Shelves for Freezer (Must have Model Number)
4015-MODEL NUMBER
background
background
background
background
background
Connection to the alarm relays and audible alarm require external control relays that are not part of the
control assembly. One relay is utilized to provide remote contacts for customer connection for remote
alarm indication (shown as CR1 in the following diagram) and is optional. The other relay is utilized for the
audible alarm silence (shown as CR2 in the following diagram) which is required.
Note that alarm 3 (low battery) is utilized for indication only. When an alarm relay is energized, the
corresponding alarm indicator, AL1 for alarm 1 (temperature), AL2 for alarm 2 (power failure) or AL3 for
alarm 3 (low battery), will be illuminated on the upper left of the C42 LCD display. When either a high/low
temperature or power failure alarm activates, the audible alarm will sound as power from the alarm
contact will pass through the normally closed contact of relay CR2. When the silence button is pressed,
the coil of relay CR2 will be energized. Since the normally open contact of CR2 will then close, power will
pass through the relay contact and will in turn hold the relay coil energized even after the push-button is
released.
This will hold the normally closed contact of the relay open, removing power from the audible alarm so
that it will no longer sound. Once both alarm conditions are no longer present, both alarm 1 and alarm 2
outputs of the C42 will be de-energized. This will in turn remove power from the coil of relay CR2 causing
it to de-energize. This will then ensure that upon the next activation of either the temperature or power
failure alarm output, power will again pass to the audible alarm causing it to sound until the alarm
condition clears or the operator presses the silence button.
WHITE
RED
BLUE
background
FDC_C-Series_User_Manual_UM0C62H_January-2021.docx Page 13 of 131
1.3 Specifications
Specification
C22
C62
C82
C83
C42
R22
Power Supply
90 to 250VAC, 47 to63Hz, 20 to 28 VAC,47-63Hz / 11 to 40 VDC
Power Consumption
C22 / R22: 8VA, 4W Maximum., C62: 10VA, 5W Maximum., C82 / C83 / C42: 12VA,6W Maximum
Over Voltage Category
II
Signal Input
Type
Thermocouple (J, K, T, E, B, R, S, N, L, U, P, C, D), RTD(PT100(DIN), PT100(JIS)), Current(mA), Voltage (V, mV)
Resolution
18 Bits
Sampling Rate
5 Times / Second (200msec)
Maximum Rating
-2VDC minimum, 12VDC maximum
Input Characteristics
Type
Range
Accuracy @ 25°C
Input Impedance
J
-120°C to 1000°C ( -184°F to 1832°F)
±2°C
2.2 MΩ
K
-200°C to 1370°C (-328°F to 2498°F)
±2°C
2.2 MΩ
T
-250°C to 400°C ( -418°F to 752°F)
±2°C
2.2 MΩ
E
-100°C to 900°C ( -148°F to 1652°F)
±2°C
2.2 MΩ
B
0°C to 1820°C (32°F to 3308°F)
±2°C (200°C to 1800°C)
2.2 MΩ
R
0°C to 1767.8°C (32°F to 3214°F)
±2°C
2.2 MΩ
S
0°C to 1767.8°C (32°F to 3214°F)
±2°C
2.2 MΩ
N
-250°C to 1300°C ( -418°F to 2372°F)
±2°C
2.2 MΩ
L
-200°C to 900°C ( -328°F to 1652°F)
±2°C
2.2 MΩ
U
-200°C to 600°C (-328°F to 1112°F)
±2°C
2.2 MΩ
P
0°C to 1395°C (32°F to 2543°F)
±2°C
2.2 MΩ
C
0°C to 2300°C (32°F to 4172°F)
±2°C
2.2 MΩ
D
0°C to 2300°C (32°F to 4172°F)
±2°C
2.2 MΩ
Land Jewel
0°C to 1880°C (32°F to 3416°F)
(Not available for C22, C62 & R22)
±2°C
2.2 MΩ
PT100(DIN)
-200°C to 850°C ( -328°F to 1562°F)
±0.4°C
1.3KΩ
PT100(JIS)
-200°C to 600°C ( -328°F to 1112°F)
±0.4°C
1.3KΩ
mA
-3mA to 27mA
±0.05%
2.5Ω
VDC
-1.3VDC to 11.5VDC
±0.05%
1.5MΩ
mV
0 to 50mV
±0.05%
2.2 MΩ
Temperature Effect
1.5µV /°C for all inputs except mA input, 3.0µV /°C for mA
Sensor Lead Resistance
Effect
Thermocouple: 0.2 µV /°Ω; 3-wire RTD: 2.6°C /Ω of Difference of Resistance of two leads
2-wire RTD: 2.6°C /Ω of Sum of Resistance of two leads
Burn-out Current
200nA
CMRR
120 dB
NMRR
55dB
Sensor Break Detection
Sensor open for Thermocouple, RTD and mV inputs, Sensor short for RTD input,
Below 1mA for 4-20mA input, Below 0.25VDC for 1 - 5VDC input, Not available for other inputs.
Sensor Break Response
Time
Within 4 seconds for Thermocouple, RTD and mV inputs, 0.1 second for 4-20mA and 1 - 5VDC inputs.
Remote Set Point Input
Type
Linear Current, Linear Voltage
Range
-3mA to 27mA, -1.3VDC to 11.5VDC
Accuracy
±0.05 %
Remote Set Point
Option
Not Available
Not Available
Available
Available
Available
Available
Not Available
Input Impedance
Current: 2.5Ω, Voltage:1.5MΩ
Resolution
18 Bits
Sampling Rate
1.66 Times/Second
Maximum Rating
280mA maximum for Current Input, 12VDC Maximum for Voltage Input
Temperature Effect
±1.5µV/°C for Voltage Input, ±3.0µV/°C for Current Input
Sensor Break
Detection
Below 1mA for 4-20mA input, Below 0.25VDC for 1 - 5VDC input, Not available for other inputs.
Sensor Break
Responding Time
0.1 Second
background

Specifications

So-Low U40-18 Questions and Answers