Version 23.00, November 2020
Rohde & Schwarz R&S
®
RTH1002, R&S
®
RTH1004 Handheld Oscilloscope 23
R&S
®
RTH-K33
Sources
R&S
®
RTH1004
CH1, CH2, CH3, CH4
R&S
®
RTH1002
CH1, CH2
Number of counters
2
Number of measurements
up to 2 using internal clock as reference;
1 using external signal to second counter
as reference
Frequency range
R&S
®
RTH1002 and R&S
®
RTH1004
10 Hz up to 66 MHz
R&S
®
RTH1002 with -B221 option and
R&S
®
RTH1004 with -B241 option
10 Hz up to 110 MHz
R&S
®
RTH1002 with -B222 option and
R&S
®
RTH1004 with -B242 option
10 Hz up to 220 MHz
R&S
®
RTH1002 with -B223 option and
R&S
®
RTH1004 with -B243 option
10 Hz up to 385 MHz
R&S
®
RTH1002 with -B224 option and
R&S
®
RTH1004 with -B244 option
10 Hz up to 550 MHz
Long term statistics
max., min. and average since (re-)start;
max. and min. are time stamped with a
resolution of 1 s.
Precision
(measurement accuracy must also take
reference clock accuracy into account)
f 10 kHz
7 digits (meas.)
1 kHz f < 10 kHz
6 digits (meas.)
100 Hz f < 1 kHz
4 digits (meas.)
10 Hz f < 100 Hz
3 digits (meas.)
(greater precision can be achieved by
observing the long term average)
Coupling
Measured channels are automatically
switched to AC coupling.