Menu Structure
Temp Setup Menu
Menu Structure4
Temp Setup Menu4.1
9142/9143/9144 MAIN MENU
F1 F2 F3 F4
TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-P only)
F1 SETUP
SCAN RATE: <edit> Degrees per minute at which the Field Metrology Well increases or decreases temperature
STABLE LIMIT: <edit> The defined minimum stability over a 2 minute period before alarm is activated
STABLE ALARM: <ENABLE, DISABLE> <select> Audible alarm indicating the Field Metrology Well is stable
F2 CUTOUT
SOFT CUTOUT: <edit> User adjustable safety cutout
HARD CUTOUT: <view only> Factory set safety cutout
Hot Keys (while viewing main screen)
SETPoinT. Key - SETPOINT
SETPOINT:
<edit> Set point temperature
ENTER <enable control of the instrument>
F1 SELECT PRESET <1-8> <select>
F1 EDIT PRESET <1-8> <edit>
F4 – SAVE/DISABLE <disables control of instrument>
ºC / ºF Key - Units: <°C, °F> Changes temperature units
Up/Down Arrow Keys <toggle> <adjust contrast>
Up Key: Darker
Down Key: Lighter
F1 & F4 Keys (same time) <reset display language to English>
F1 & F3 Keys (same time) <enable/disable key press beep>
1 Beep – Valid key entry
2 Beep – Invalid key entry
Code Update Mode Keys
ENTER & EXIT Keys (hold during power up) <initiate code update mode> Allows instrument software to be updated
Figure 11 Main Menu - Temp SetUp