Loading ...
Loading ...
Loading ...
15
Specications and Environmental Conditions
Environmental Conditions
-P Specifications
Built-in TC Thermometer Readout Accuracy Type J: ± 0.7 °C at 660 °C
Type K: ± 0.8 °C at 660 °C
Type T: ± 0.8 °C at 400 °C
Type E: ± 0.7 °C at 660 °C
Type R: ± 1.4 °C at 660 °C
Type S: ± 1.5 °C at 660 °C
Type M: ± 0.6 °C at 660 °C
Type L: ± 0.7 °C at 660 °C
Type U: ± 0.75 °C at 600 °C
Type N: ± 0.9 °C at 660 °C
Type C: ± 1.1 °C at 660 °C
TC Millivolt Range –10 mV to 75 mV
Voltage Accuracy 0.025 % of reading +0.01mV
Internal Cold Junction Compensation Accuracy ± 0.35 °C (ambient of 13 °C to 33 °C)
TC Connection Small connectors
Built-in mA Readout Accuracy 0.02% of reading + 0.002 mA
mA Range Cal 4-22 mA, Spec 4-24 mA
mA Connection 2 terminal input
Loop Power Function 24 VDC loop power
Built-in Electronics Temperature Coefficient
(0 °C to 13 °C, 33 °C to 50 °C)
± 0.005 % of range per °C
The temperature range may be limited by the reference probe connected to the readout. The Built-
In Reference Accuracy does not include the sensor probe accuracy. It does not include the probe
uncertainty or probe characterization errors.
Measurement accuracy specications apply within the operating range and assume 4-wires for PRTs.
With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference
between the resistances of the lead wires.
Environmental Conditions2.2
Although the instrument has been designed for optimum durability and trouble-free operation,
it must be handled with care. The instrument should not be operated in an excessively dusty or
dirty environment. Maintenance and cleaning recommendations can be found in the Mainte-
nance section. The instrument operates safely under the following environmental conditions:
ambient temperature range: 0-50°C (32-122°F)
ambient relative humidity: 0 % to 90 % (non-condensing)
mains voltage: within ±10% of nominal
vibrations in the calibration environment should be minimized
altitude: less than 2,000 meters
indoor use only
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...