1587 FC/1587/1577
Users Manual
42
Model 1587T
Output Voltage Display Range Resolution Test Current
Resistance Accuracy
±(% of Rdg + Counts)
50 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 50 kΩ ±(3 % + 5 counts)
6.0 to 50.0 MΩ 0.1 MΩ
100 V
(0 % to +20 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 100 kΩ ±(3 % + 5 counts)
6.0 to 60.0 MΩ 0.1 MΩ
60 to 100 MΩ 1 MΩ
Model 1587C FC
Output Voltage Display Range Resolution Test Current
Resistance Accuracy
±(% of Rdg + Counts)
50 V
(-10 % to +10 %)
0.01 to 6.00 MΩ 0.01 MΩ
1 mA @ 50 kΩ ±(3 % + 5 counts)
6.0 to 50.0 MΩ 0.1 MΩ
100 V
(-10 % to +10 %)
0.01 to 6.00 MΩ 0.01 MΩ 1 mA @ 100 kΩ ±(3 % + 5 counts)
250 V
(-10 % to +10 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 250 kΩ ±(1.5 % + 5 counts)
60 to 250 MΩ 1 MΩ
500 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 500 kΩ ±(1.5 % + 5 counts)
60 to 500 MΩ 1 MΩ
1000 V
(0 % to +20 %)
0.1 to 60.0 MΩ 0.1 MΩ
1 mA @ 1 MΩ
±(1.5 % + 5 counts)
60 to 600 MΩ 1 MΩ
0.6 to 2.0 GΩ 100 MΩ ±(10 % + 3 counts)
1.888.610.7664 sales@GlobalTestSupply.com