3. Set the channel to ON if necessary.
4. Push to open the Copy Channel menu.
5. Use or to highlight a channel then push to select the channels to paste to.
6. Push to paste the channels.
Save or Load a Channel Configuration (Setup File)
The channel and test configuration can be saved to a file referred to as the Setup file.
When the Setup file is saved, the Product saves the present configuration of each channel
(including Mx+B and Alarms) in their current state (ON or OFF). Setup files can be
loaded at a later time or transferred to another Product with a USB drive. Setup files are
managed and loaded from the internal memory in the Memory menu.
Notes
Setup files cannot be directly saved to or loaded from the USB drive.
To save Setup files to a USB drive, save the file to the internal
memory, then copy it to the USB drive. To load a Setup file from a
USB drive, copy the file to internal memory then load the file. See the
instructions below.
The Product will not load a Setup file if the present configuration does
not match the configuration in the Setup file. This is caused by missing
Input Modules. Match the configuration and try to load the Setup file
again.
To save a Setup File:
1. Push
.
2. Push .
3. Follow the on-screen directions to name the file then push to save.
To save a Setup file to a USB drive:
1. Save the current Setup file to internal memory.
2. Insert a USB drive into the front-panel USB port.
Note
Wait about 15 seconds to let the Product recognize the USB drive. The LED
illuminates solid red when the USB drive is recognized and ready.
3. Push .
4. Push to select
Internal Files.