753/754
Users Manual
80
8. Push the applicable function key for the control output
(for example, or ). If the transmitter is connected
to a current loop that has a power supply, select
Simulate Transmitter for the current output.
9. Select a source value, for example, 4 mA.
10. Push for MEASURE/SOURCE mode.
11. Push More Choices until the Transmitter Mode
softkey is shown.
12. Push the Transmitter Mode softkey.
13. Set the 0 % and 100 % values for MEASURE and
SOURCE on the display. Linear or can be selected
for the transfer function.
14. Push the Done softkey.
The Product is now in Transmitter mode. It measures
the process input and sources the control signal output
proportional to the input.
15. To change the Transmitter mode parameters, push
Change Setup, and do the procedure in step 13 again.
16. To exit Transmitter mode, push the Abort softkey.
Memory Operations
Save Results
As Found/As Left test results are automatically kept at the
end of each test routine. Any other time during MEASURE,
SOURCE, or MEASURE/SOURCE, if necessary, push the
Save softkey to keep the data on the display for later
inspection.