109
DC Current Measurement
Range Resolution
% of Reading + Floor
1 Year 2 Year
±30.000 mA 1 μA
0.01 % + 5 μA 0.015 % + 7 μA
±100.00 mA 10 μA
0.01 % + 20 μA 0.015 % + 30 μA
Maximum Input: 110 mA
Maximum Burden Voltage: 420 mV at 22 mA
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
No Fuse
Normal mode rejection: 90 dB at 50 or 60 Hz nominal, and 60 dB at 1200 Hz and 2200 Hz nominal (HART signals)
Resistance Measurement
Range Resolution
% of Reading + Floor
Source
Current
1 Year 2 Year
10.000 Ω 0.001 Ω 0.05 % + 0.050 Ω 0.07 % + 0.070 Ω
3 mA
100. 00 Ω 0.01 Ω 0.05 % + 0.05 Ω 0.07 % + 0.07 Ω
1 mA
1.0000 kΩ 0.1 Ω 0.05 % + 0.0005 kΩ 0.07 % + 0.0007 kΩ 500 μA
10.000 kΩ 1 Ω 0.10 % + 0.010 kΩ 0.15 % + 0.015 kΩ 50 μA
Open-circuit voltage: 5 V nominal
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Continuity Testing
Tone Resistance
Continuous tone
May or may not get tone
No tone
<25 Ω
25 to 400 Ω
>400 Ω