Obsolete
PF_CONTROL | PFCT
Command/Query
DESCRIPTION
The PF_CONTROL command controls the
pass/fail controlling options: ―operate‖, ―output‖
and the ―stop on output‖.
See instrument‘s Operator Manual for these
options.
The PF_CONTROL? query returns the
controlling options of the pass/fail.
COMMAND SYNTAX
PF_CONTROL
TRACE,<trace>,CONTROL,<control>,OUTPU
T,<output>,OUTPUTSTOP,<state>
<trace>:={C1,C2,C3,C4}
<control>:={START,STOP}
<output>:={FAIL,PASS}
<state>:={ON,OFF}
QUERY SYNTAX
PF_ CONTROL?
RESPONSE FORMAT
PF_CONTROL
TRACE,<trace>,CONTROL,<control>,OUTPU
T,<output>,OUTPUTSTOP,<state>
EXAMPLE
The following command sets source to Channel
1, operate to start, output to pass and
stop on output to off.
Command message:
PFCT
TRACE,C1,CONTROL,START,OUTPUT,PASS,
OUTPUTSTOP,OFF
RELATED COMMANDS
PFSC
PFBF
PFOP
PFFS
Note: