Siglent SDS1204X-E oscilloscope

DSO programming guide - Page 120

For SDS1204X-E.

PDF File Manual, 256 pages, Read Online | Download pdf file

SDS1204X-E photo
Loading ...
Loading ...
Loading ...
Digital Oscilloscope Series
120
PASS/FAIL
PF_SOURCE | PFSC
Command/Query
DESCRIPTION
The PF_SOURCE command sets measurement
sources for Pass/Fail test.
The PF_SOURCE? query returns the
measurement source for Pass/Fail test.
COMMAND SYNTAX
PF_SOURCE <trace>
<trace>:={C1,C2,C3,C4}
QUERY SYNTAX
PF_SOURCE?
RESPONSE FORMAT
PF_SOURCE <trace>
EXAMPLE
The following command sets the measurement
source to Channel 1 when Channel 1 is on.
Command message:
PFSC C1
Note:
The table below shows the availability of command in each oscilloscope series.
Model
Valid?
SDS1000CFL
no
SDS1000A
no
SDS1000CML+/CNL+/DL+/E+/F+
no
SDS2000X
no
SDS1000X
no
SDS1000X-E
yes
Loading ...
Loading ...
Loading ...