Fast Auto/Manual Range Toggle
When NOT in a menu mode, push for ½ second to toggle between
Auto Range and Manual Range.
Fast Auto Rescale
When in Manual Range and NOT in a menu mode, push for ½ second
to automatically rescale the level and span range for objects in the thermal
field of view. This feature operates the Imager in a semi-automatic mode if
manual fine re-adjustment of level and span with the arrow buttons is not
necessary. Rescaling can be done as often, or as little, as needed.
Note
The Imager always powers up in the same Range mode, Auto or
Manual, as when it was powered down.
Level for Manual Operation Mode
When put into manual ranging, the level setting moves the thermal span up or
down within the total temperature range. See Figure 4. In the live manual
mode, the arrow buttons are always available to adjust the level and span.
To set the level:
1. Push to move the range to a higher temperature level.
2. Push to move the range to a lower temperature level.
While you adjust the manual level, the scale along the right side of the display
shows the thermal span as it moves to different levels within the total range.