Basic Operation
Turn On and Off the Product
Before you use the Product for the first time, charge the battery. See Charge Batteries.
To turn on the Product, push and hold O for 1 second. To turn off the Product, push and
hold O for 2 seconds.
Note
All thermal Imagers need sufficient warm-up time for accurate temperature
measurements and best image quality. Warm-up time can vary by model and
environmental conditions. Although most Imagers are fully warmed up in 3 minutes
to 5 minutes, wait a minimum of 10 minutes if the most accurate temperature
measurement is important to your application. When you move an Imager between
environments with large differences in ambient temperature, allow for additional
adjustment time.
Focus
Correct focus makes sure that the infrared energy is correctly directed onto the pixels of the
detector. Without correct focus, the thermal image can be blurry and the radiometric data
may be inaccurate. Out-of-focus infrared images are frequently unusable or of little value.
To focus with the advanced manual focus system, rotate the Manual Focus Control until the
inspection object is in proper focus.
To simulate fixed focus at 1.2 m, align the raised bar on the focus control with the marker on
the case. See Figure 1.
Figure 1. Fixed Focus
1.2 m