FLUKE 9009-B-156 Dual Range Dry-Block Calibrator, Black

User Manual for the Fluke 9009-B-156 Dual Range Dry-Block Calibrator, Black - Page 36

For 9009-B-156.

PDF File Manual, 45 pages, Read Online | Download pdf file

9009-B-156 photo
Loading ...
Loading ...
Loading ...
9009 Industrial Dual-Block Calibrator
User’s Guide
32
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.00 C
Set current set-point to n s[etpoint]=n s=350 Instrument
Range
Read temperature t[emperature] t t{h or c}: 9999.99 {C or
F}
th: 55.66 C
Read temperature units u[nits] u u: x u: C
Set temperature units: u[nits]=c/f
CorF
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
u[nits]=f u=f
Read scan mode sc[an] sc sc: {ON or OFF} sc: ON
Set scan mode sc[an]=on/off sc=on ON or OFF
Read scan rate sr[ate] sr srat: 99.9 {C or F}/min srat:12.4 C/min
Set scan rate sr[ate]=n sr=1.1 0.1 to 99.9
Secondary Menu
Read proportional band setting pr[opband] pr pb: 999.9 pb: 15.9
Set proportional band to
n
pr[opband]=n pr=8.83 Depends on
Configuration
Read heater power
(duty cycle)
po[wer] po po: 999.9 po: 1.3
Configuration Menu
Operating Parameters Menu
Read High Limit hl hl hl: 999 hl: 350
Set High Limit hl=
n
hl=350 Cold 25 to 126,
Hot50to350
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=n sa=0 0 to 999
Set serial duplex mode: du[plex]=f[ull]/h[alf]
FULL or HALF
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Table 2 9009 Controller Communications Commands
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...