3 Specifications and Environmental
Conditions
3.1 Specifications
Specification Hot Block Cold Block
Range
50°C to 350°C (122°F to 662°F) –15°C to 110°C (5°F to 230°F)
[–8°C (17°F) to 110°C (230°F) when
hot block is at 350°C(662°F)]
Accuracy (Ambient at 23°C
±5°C)
±0.6°C ±0.2°C
Stability
±0.1°C from 50°C to 100°C
±0.05°C > 100°C
±0.05°C
Well-to-Well Uniformity
±0.1°C
Display Resolution
0.1°
Heating Times
30 minutes from 25°C to 350°C 15 minutes from 25°C to 110°C
Cooling Times
40 minutes from 350°C to 100°C 16 minutes from 25°C to -15°C
Stabilization Times
8 minutes
Well Depth
4 inches(102 mm)
Removable Inserts
Two 1/4”(6.4mm) and 3/16”(4.8mm)inserts included – other inserts available
Power
115 VAC (±10%), 50/60 Hz, 3 A, 250 W
[optionally 230 VAC (±10%), 50/60 Hz, 1.6 A, 250 W]
Heater
135 W 40 W TED
Cooling
Fan 40 W TED
Size
7”H x 10.5”W x 9.75”D (178 x 267 x 248 mm)
Weight
10 lb (4.5 kg)
Safety
OVER VOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per
IEC1010-1
Stability is two times the standard deviation.
Heating and cooling times may be affected by line voltages and ambient temperatures.
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou
-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
ambient temperature range: 5 - 45°C (41 - 113°F)
9
3 Specifications and Environmental Conditions
Specifications
1.888.610.7664 sales@GlobalTestSupply.com