Pressure Calibrator
Specifications
27
Product Compliance Markings
E
( II 1 G EEx ia IIC T4
Kema 04ATEX1061 X
F
Class I Div. 1 Groups A-D T4
AEx ia IIC T4
Ta = -10 °C +55 °C
; Conforms to relevant Australian standards.
Manufactured by Martel Electronics Inc., 1F Commons Drive, Londonderry, NH USA
Additional Safety Information: Complies with CAN/CSA C22.2 No. 1010.2:1995. Complies with ANSI/ISA S82.01-1995.
Complies with IEC 61010-1-95 CAT I, 30 V
Entity Parameters:
Vi, Ui Ii Pi Ci Li
30 V 250 mA 1.88 W 0 μF 0 mH
Co Lo
Vo, Uo Io Po
IIC IIB IIA IIC IIB IIA
7.14 V 1.09 mA 1.9 mW 13.5 μF 240 μF 1000 μF 1.0 H 3.0 H 8.0 H
Power requirements: See Approved Batteries.
Size: 66 mm H x 94 mm W x 216 mm L (2.60 in H x 3.70 in W x 8.5 in L)
Weight: 992 g (35 oz)