FLUKE 718EX-30 G Intrinsically Safe Pressure Calibrator

User Manual for the Fluke 718Ex-30G Intrinsically Safe Pressure Calibrator - Page 14

For 718EX-30.

PDF File Manual, 33 pages, Read Online | Download pdf file

718EX-30 photo
Loading ...
Loading ...
Loading ...
718Ex 30G/100G/300G
Users Manual
8
The safety features and integrity of the unit may be
compromised by any of the following:
External damage to the housing
Internal damage to the Calibrator
Exposure to excessive loads
Incorrect storage of the unit
Damage sustained in transit
Correct certification is illegible
Using the product with the red holster removed
Functioning errors occur
Permitted limitations are exceeded
Functioning errors or obvious measurement
inaccuracies occur which prevent further
measurement by the Calibrator
Opening the case
Safety Regulations
The use of the Calibrator meets the requirements of the
regulations providing that the user observes and applies
the requirements as stated in the regulations and that
improper and incorrect use of the unit is avoided.
Use must be restricted to the specified application
parameters.
Do not open the Calibrator.
Do not remove or install the battery within the
Ex-hazardous area.
Do not carry additional batteries within the
Ex-hazardous area.
Use only type-tested batteries. The use of any other
batteries will invalidate the Ex-certification and present
a safety risk.
Do not use the Calibrator in an Ex-hazardous area
unless it is completely and securely fitted in its
accompanying red holster.
Only use the Calibrator in circuits with compatible
entity parameters.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct
.com
Loading ...
Loading ...
Loading ...