FLUKE 718EX-30 G Intrinsically Safe Pressure Calibrator

User Manual for the Fluke 718Ex-30G Intrinsically Safe Pressure Calibrator - Page 27

For 718EX-30.

PDF File Manual, 33 pages, Read Online | Download pdf file

718EX-30 photo
Loading ...
Loading ...
Loading ...
Pressure Calibrator
Maintenance
21
3. Press
S
to enter pressure switch test mode. The
Calibrator will display CLOSE instead of a mA
measurement.
4. Apply pressure with the pump slowly until the switch
opens.
Note
In the switch test mode, the display update rate is
increased to help capture changing pressure
inputs. Even with this enhanced sample rate,
pressuring the device under test should be done
slowly to ensure accurate readings.
5. OPEN is displayed once the switch is open. Bleed the
pump slowly until the pressure switch closes. RCL
appears on the display.
6. Press
S to read the pressure values for when the
switch opened, for when it closed, and for the
deadband.
7. Hold
Sfor three seconds to exit the switch test or
press
c to reset the switch test.
Maintenance
XW Warning
To avoid possible electric shock, personal
injury, or sudden release of pressure, review
Safety Information before proceeding.
For maintenance procedures not described in this manual,
or if the Calibrator needs repair, contact a Fluke Service
Center. See Contacting Fluke.
In Case of Difficulty
After removing the Calibrator from the Ex-hazardous
area, check the battery, test leads, pressure module,
and pressure tubing. Follow replacement and
connection instructions properly.
Review this manual and control drawing to make sure
the Calibrator is used correctly.
If the Calibrator needs repair, and the Calibrator is under
warranty, see the warranty statement for terms. If the
warranty has lapsed, the Calibrator can be repaired and
returned for a fixed fee.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct
.com