715 English Instruction Sheet
Page 4
1. With the Calibrator OFF, press the
O key. P.S.xx is
displayed, where xx is the turn-off time in minutes. OFF
means the power saver is disabled.
2. Press
] and/or } to increase or decrease the turn-
off time in minutes.
3. To disable, press } until the display shows OFF.
HART
TM
Resistor Mode
The calibrator has a user selectable 250 ohm HART
TM
resistor to
facilitate use with HART
TM
communication devices. The resistor
can be switched in/out at any time by simultaneously pressing
the
V and A keys. Use a HART
TM
communicator when
measuring dc mA with loop power or sourcing mA.
Span Check Feature
The calibrator allows the users to store 0% and 100% setpoints
for each output function. Once setpoints are stored, the span
check feature allows the user to quickly toggle back and forth
from 0% to 100% or to step in 25% increments.
Automatic step and ramp modes can be enabled while in the
span check mode by simultaneously pressing the
] } keys.
First select the desired output mode (V,mV or mA) then preceed
to store the setpoints:
1. Storing 0% and 100% Setpoints
2. Using the ] } controls set the output to the desired
output value for 0%.
3. Press the ] } scroll keys simultaneously to store
the 0% value.
4. Now using the
] } controls set the output to the
desired output value for 100%
5. Again, press the ] } scroll keys simultaneously to
store the 100% value.