FLUKE 715 Volt/mA Calibrator

Instruction Sheet - Page 14

For 715.

PDF File Manual, 16 pages, Read Online | Download pdf file

715 photo
Loading ...
Loading ...
Loading ...
715 English Instruction Sheet
Page 14
Specifications
Specifications are based on a one year calibration cycle and
apply from +18°C to +28°C unless stated otherwise. “Counts”
means number of increments or decrements of the least
significant digit.
DC V Input and Output
Range Resolution Accuracy,
±(% of Reading + Counts)
200 mV 0.01 mV 0.015% + 2
20 V output
25 V input
0.001 V 0.01% + 2
Input impedance: 1 M
(nominal), < 100 pF
Overvoltage protection: Fuseless
Voltage drive capability: 1 mA
DC mA Input
Range Resolution Accuracy,
±(% of Reading + Counts)
24 mA 0.001 mA 0.01% + 2
Overload protection: Fuseless
DC mA Output
Range: 0 mA to 24 mA
Percent display: 0% = 4 mA, 100% = 20 mA
Accuracy: ±(0.01% of reading + 2 counts)
Source mode:
Compliance: 1000 at 20 mA for battery voltage 6.8 V (700
at 20 mA for battery voltage 5.8 to 6.8 V)
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum,
12 V minimum
Loop Power
24 V ±10%
1.888.610.7664 sales@GlobalTestSupply.com