FLUKE 721EX-1615 Dual Sensor Pressure Calibrator, 16 PSIG/1500 PSIG

User Manual for the Fluke 721EX-1615 Dual Sensor Pressure Calibrator, 16 PSIG/1500 PSIG - Page 35

For 721EX-1615.

PDF File Manual, 46 pages, Read Online | Download pdf file

721EX-1615 photo
Loading ...
Loading ...
Loading ...
Pressure Calibrator
Transmitter Calibration
27
Transmitter Calibration
mA Input Function
The mA input function reads back the 4 mA to 20 mA
output from the device being calibrated. This can be done
passively. The device under test directly generates 4 mA
to 20 mA and can be read by the Product.
hvf047.eps
Pressure-to-Current Transmitter Calibration
To calibrate a pressure-to-current transmitter (P/I):
1. Connect the Product and the pump to the transmitter.
See Figure 7.
2. Apply pressure with the pump.
3. Measure the current output of the transmitter.
4. Ensure the reading is correct. If not, adjust the
transmitter as necessary.
Note
Use low-volume tubing when possible.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com