FLUKE 721EX-1615 Dual Sensor Pressure Calibrator, 16 PSIG/1500 PSIG

User Manual for the Fluke 721EX-1615 Dual Sensor Pressure Calibrator, 16 PSIG/1500 PSIG - Page 25

For 721EX-1615.

PDF File Manual, 46 pages, Read Online | Download pdf file

721EX-1615 photo
Loading ...
Loading ...
Loading ...
Pressure Calibrator
Product Features
17
Example:
A pressure transmitter under test is 30 psi (2 Bar) Full-
Scale and outputs a corresponding 4 mA to 20 mA signal.
Program a 0 psi to 30 psi pressure span into the Product
and the Product calculates and shows the deviation or
%Error from the expected 4 mA to 20 mA output. This
eliminates the need for manual calculations and helps
when it is difficult to set an exact pressure.
To use the %ERROR function:
1. Push to access the menus.
2. Push to step through the menu until %ERROR is
shown on the menu bar.
3. Push to open the %ERROR screen.
4. Push to scroll through the port choices (P1 or P2).
5. Push to configure the pressure range settings.
6. Use the arrow keys to set the 100 % point of the
desired pressure range, select DONE SET when
finished.
7. Use the arrow keys to set 0 % point and select
DONE SET. The % ERROR mode is now ready to
use.
Note
The 0 % and 100 % point will be saved in non-
volatile memory until they are changed again by
the user.
hmq054.eps