9
Specications and Environmental Conditions
Specications
Specications and Environmental Conditions3
3.1 Specications
Table 2 Specifications
Range -30 to 125°C (-22 to 257°F)
Accuracy ±0.25°C
Stability ±0.03°C at -25°C (oil, 5010)
±0.05°C at 125°C (oil, 5010)
Uniformity ±0.02°C
Resolution 0.01°C/F
Operating Temperature 5 to 45°C (41 to 113°F)
Heating Time 25°C to 100°C: 35 minutes
Cooling Time 25°C to –25°C: 45 minutes
Well Size 2.5" dia. x 5.5" deep (64 x 139 mm)
(access opening is 1.9" [48 mm] in diameter)
Size 9" W x 13.2" H x 10.5" D (23 x 34 x 26 cm)
Weight 22 lb. (10 kg) with uid
Display LED, Switchable °C or °F
Cooling Fan and Thermal Electric Devices (TED)
Fault Protection Sensor burnout and short protection
Power 94 – 234 VAC (±10%), 50/60 Hz, 400 W
Safety Overvoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01
Fuse Rating 250 V 5 A F(fast acting) NO USER SERVICEABLE PARTS
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and clean ing recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
ambient temperature range: 5 – 45°C (41 – 113°F)
ambient relative humidity: maximum 80% for temperature <31°C, de creasing
linearly to 50% at 40°C
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters
1.888.610.7664 sales@GlobalTestSupply.com